dorsal/arxiv
View SchemaLeakage Tests for Am-241 Solid Sources Used for Liquid Xenon Detector Monitoring
| Authors | S. Bazzarri, G. Cicoli, P. De Felice, G. Rossi, F. Sedda |
|---|---|
| Categories | |
| ArXiv ID | physics/0509101 |
| URL | https://arxiv.org/abs/physics/0509101 |
| DOI | 10.1016/j.nima.2006.02.170 |
Abstract
Experimental tests are described to check solid Am-241 radioactive sources used to monitor operation of liquid xenon detectors. In particular, radioactive leakage was tested in extreme temperature conditions following immersion in liquid nitrogen for different time periods. No radioactivity loss was detected from the sources considered. The paper describes the source characteristics, test methods and results obtained.
{
"annotation_id": "a32ef57e-4753-476c-9c92-5410180c12d3",
"date_created": "2026-03-02T18:00:59.936000Z",
"date_modified": "2026-03-02T18:00:59.936000Z",
"file_hash": "353b1669655319c2a62b600a7332c8bd0c339d319f1f4948ec10ee82a21f5c3a",
"private": false,
"record": {
"abstract": "Experimental tests are described to check solid Am-241 radioactive sources\nused to monitor operation of liquid xenon detectors. In particular, radioactive\nleakage was tested in extreme temperature conditions following immersion in\nliquid nitrogen for different time periods. No radioactivity loss was detected\nfrom the sources considered. The paper describes the source characteristics,\ntest methods and results obtained.",
"arxiv_id": "physics/0509101",
"authors": [
"S. Bazzarri",
"G. Cicoli",
"P. De Felice",
"G. Rossi",
"F. Sedda"
],
"categories": [
"physics.med-ph",
"physics.ins-det"
],
"doi": "10.1016/j.nima.2006.02.170",
"title": "Leakage Tests for Am-241 Solid Sources Used for Liquid Xenon Detector Monitoring",
"url": "https://arxiv.org/abs/physics/0509101"
},
"schema_id": "dorsal/arxiv",
"source": {
"execution_id": "caa3dd26-8277-4468-a5dc-f6ea4bada4af",
"id": "arXiv Dataset IDs",
"type": "Model",
"variant": "snapshot-2026-03-01",
"version": "0.1.0"
},
"user_id": 1000002
}