View SchemaPlanar LAAPDs: Temperature Dependence, Performance, and Application in Low Energy X-ray Spectroscopy
| Authors | L. Ludhova, F. D. Amaro, A. Antognini, F. Biraben, J. M. R. Cardoso, C. A. N. Conde, D. S. Covita, A. Dax, S. Dhawan, L. M. P. Fernandes, T. W. Hansch, V. W. Hughes, O. Huot, P. Indelicato, L. Julien, P. E. Knowles, F. Kottmann, J. A. M. Lopes, Y. -W. Liu, C. M. B. Monteiro, F. Mulhauser, F. Nez, R. Pohl, P. Rabinowitz, J. M. F. dos Santos, L. A. Schaller, D. Taqqu, J. F. C. A. Veloso |
|---|---|
| Categories | |
| ArXiv ID | physics/0410099 |
| URL | https://arxiv.org/abs/physics/0410099 |
| DOI | 10.1016/j.nima.2004.11.017 |
| Journal | Nucl.Instrum.Meth. A540 (2005) 169-179 |
Abstract
An experiment measuring the 2S Lamb shift in muonic hydrogen mup is being performed at the Paul Scherrer Institute, Switzerland. It requires small and compact detectors for 1.9 keV x rays (2P-1S transition) with an energy resolution around 25% at 2 keV, a time resolution better than 100 ns, a large solid angle coverage, and insensitivity to a 5 T magnetic field. We have chosen Large Area Avalanche Photodiodes (LAAPDs) from Radiation Monitoring Devices as x-ray detectors, and they were used during the last data taking period in 2003. For x-ray spectroscopy applications, these LAAPDs have to be cooled in order to suppress the dark current noise, hence, a series of tests were performed to choose the optimal operation temperature. Specifically, the temperature dependence of gain, energy resolution, dark current, excess noise factor, and detector response linearity was studied. Finally, details of the LAAPDs application in the muonic hydrogen experiment as well as their response to alpha particles are presented.
{
"annotation_id": "62c36f50-73b3-4f44-a998-7361c7370a86",
"date_created": "2026-03-02T18:00:53.162000Z",
"date_modified": "2026-03-02T18:00:53.162000Z",
"file_hash": "97e43f4b85e160a8d759a258e97fe5159049329b1488d98fdc35ae9badbc11f5",
"private": false,
"record": {
"abstract": "An experiment measuring the 2S Lamb shift in muonic hydrogen mup is being\nperformed at the Paul Scherrer Institute, Switzerland. It requires small and\ncompact detectors for 1.9 keV x rays (2P-1S transition) with an energy\nresolution around 25% at 2 keV, a time resolution better than 100 ns, a large\nsolid angle coverage, and insensitivity to a 5 T magnetic field. We have chosen\nLarge Area Avalanche Photodiodes (LAAPDs) from Radiation Monitoring Devices as\nx-ray detectors, and they were used during the last data taking period in 2003.\nFor x-ray spectroscopy applications, these LAAPDs have to be cooled in order to\nsuppress the dark current noise, hence, a series of tests were performed to\nchoose the optimal operation temperature. Specifically, the temperature\ndependence of gain, energy resolution, dark current, excess noise factor, and\ndetector response linearity was studied. Finally, details of the LAAPDs\napplication in the muonic hydrogen experiment as well as their response to\nalpha particles are presented.",
"arxiv_id": "physics/0410099",
"authors": [
"L. Ludhova",
"F. D. Amaro",
"A. Antognini",
"F. Biraben",
"J. M. R. Cardoso",
"C. A. N. Conde",
"D. S. Covita",
"A. Dax",
"S. Dhawan",
"L. M. P. Fernandes",
"T. W. Hansch",
"V. W. Hughes",
"O. Huot",
"P. Indelicato",
"L. Julien",
"P. E. Knowles",
"F. Kottmann",
"J. A. M. Lopes",
"Y. -W. Liu",
"C. M. B. Monteiro",
"F. Mulhauser",
"F. Nez",
"R. Pohl",
"P. Rabinowitz",
"J. M. F. dos Santos",
"L. A. Schaller",
"D. Taqqu",
"J. F. C. A. Veloso"
],
"categories": [
"physics.ins-det"
],
"doi": "10.1016/j.nima.2004.11.017",
"journal_ref": "Nucl.Instrum.Meth. A540 (2005) 169-179",
"title": "Planar LAAPDs: Temperature Dependence, Performance, and Application in Low Energy X-ray Spectroscopy",
"url": "https://arxiv.org/abs/physics/0410099"
},
"schema_id": "dorsal/arxiv",
"source": {
"execution_id": "8c3f470a-87b7-4c91-9a1d-db5c5b59e3f4",
"id": "arXiv Dataset IDs",
"type": "Model",
"variant": "snapshot-2026-03-01",
"version": "0.1.0"
},
"user_id": 1000002
}