View SchemaComparison of discharge lamp and laser pumped cesium magnetometers
| Authors | S. Groeger, A. S. Pazgalev, A. Weis |
|---|---|
| Categories | |
| ArXiv ID | physics/0412011 |
| URL | https://arxiv.org/abs/physics/0412011 |
| DOI | 10.1007/s00340-005-1773-x |
Abstract
We have performed a comparison of laser (LsOPM) and lamp (LpOPM) pumped cesium vapor magnetometers. Although the LsOPM operated 50% above its shot-noise limit we found an intrinsic sensitivity of 15 fT/\sqrt{Hz} and 25 fT/\sqrt{Hz} for the LsOPM and LpOPM respectively. Two modes of operation, viz.,the phase-stabilized and the self-oscillating mode were investigated and found to yield a similar performance. We have compared the performance of the LsOPM and the LpOPM directly by simultaneous measurements of field fluctuations of a 2 microtesla magnetic field inside a multilayer magnetic shield and have used one of the magnetometers for an active field stabilization. In the stabilized mode we found a gradient instability of 25 fT within an integration time of 100 s, which represents an upper limit of the long-term stability of the magnetometers.
{
"annotation_id": "46c90a97-1ab6-45aa-98cf-e914a0ee5821",
"date_created": "2026-03-02T18:00:53.497000Z",
"date_modified": "2026-03-02T18:00:53.497000Z",
"file_hash": "4b1d2625a4935f5467a12fe68cda3c3a8f75995ae0ab0fbc110f37d759562131",
"private": false,
"record": {
"abstract": "We have performed a comparison of laser (LsOPM) and lamp (LpOPM) pumped\ncesium vapor magnetometers. Although the LsOPM operated 50% above its\nshot-noise limit we found an intrinsic sensitivity of 15 fT/\\sqrt{Hz} and 25\nfT/\\sqrt{Hz} for the LsOPM and LpOPM respectively. Two modes of operation,\nviz.,the phase-stabilized and the self-oscillating mode were investigated and\nfound to yield a similar performance. We have compared the performance of the\nLsOPM and the LpOPM directly by simultaneous measurements of field fluctuations\nof a 2 microtesla magnetic field inside a multilayer magnetic shield and have\nused one of the magnetometers for an active field stabilization. In the\nstabilized mode we found a gradient instability of 25 fT within an integration\ntime of 100 s, which represents an upper limit of the long-term stability of\nthe magnetometers.",
"arxiv_id": "physics/0412011",
"authors": [
"S. Groeger",
"A. S. Pazgalev",
"A. Weis"
],
"categories": [
"physics.atom-ph"
],
"doi": "10.1007/s00340-005-1773-x",
"title": "Comparison of discharge lamp and laser pumped cesium magnetometers",
"url": "https://arxiv.org/abs/physics/0412011"
},
"schema_id": "dorsal/arxiv",
"source": {
"execution_id": "8388dafd-9586-41eb-9116-e11e6858ee57",
"id": "arXiv Dataset IDs",
"type": "Model",
"variant": "snapshot-2026-03-01",
"version": "0.1.0"
},
"user_id": 1000002
}